RENEWAL PARTS
REV 4
6–2–5
Tilt Assy(FRU No. 104)
Time Required
8 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–11.
1.
Turn OFF the system.
2.
Remove the Monitor Assy (FRU No. 100) from the Swing Arm Assy. Refer to 6–2–1.
3.
Remove the Monitor Rear Cover. Refer to 6–2–4.
4.
Unscrew four screws (1 – 4).
5.
Remove the Tilt Assy.
104 Tilt Assy
1
2
3
4
TILT ASSY DISASSEMBLY
ILLUSTRATION 6–11