PERIODIC MAINTENANCE
7–5
REV 3
7–3
ELECTRICAL SAFETY TESTS
The following tests are performed at the factory and should be performed at the site. These tests are: grounding
continuity, chassis leakage current, probe leakage current, and ECG leakage current. All measurements are
made with an Electrical Safety Analyzer Model Dale 600/600E built by Dale Technology Corporation or equivalent
device.
Recommended Tool
PART NUMBER
NAME
DESCRIPTION
46–285652G1
Electric Safety Analyzer DALE 600
For 110/115V Units
46–328406G2
Electric Safety Analyzer DALE 600E
For 220/240V Units
2113015
Leakage Current Ultrasound Kit
For 100/115V and 220/240V Units
7–3–1
Outlet Test Wiring Arrangement
Test all outlets in the area for proper wiring arrangement by plugging in the Dale 600/600E and noting the
combination of lights that are illuminated (Refer to ILLUSTRATION 7–1).
AC outlet
Combination
of
Light
Condition
CORRECT
WIRING
REVERSE
POLARITY
OPEN
GROUND
OPEN
HOT
HOT/GROUND
REVERSED
OUTLET TEST
ILLUSTRATION 7–1
Note
No outlet tester can detect the condition where the Neutral (grounded supply) conductor and the
Grounding (protective earth) conductor are reversed. If later tests indicate high leakage currents,
this should be suspected as a possible cause and the outlet wiring should be visually inspected.