2–1–4
Recommended Ultrasound Room Layout
TABLE 2– 1 provides the requirements for an ultrasound room:
TABLE 2– 1
ULTRASOUND ROOM REQUIREMENTS
CURRENT RATING
2A (115V, 100V) ; 1A (220–240V) CIRCUIT BREAKER
RADIATION SHIELDING
NONE REQUIRED for ULTRASOUND ENERGY
TEMPERATURE
20–26
C (68–79
F) for PATIENT COMFORT
HUMIDITY
50% to 70% for PATIENT COMFORT
HEAT DISSIPATION
2000 BTU/Hr for LOGIQ
α
200
FLOOR LOADING
Approximately 240 – 300 kg/m
2
without Accessories
FLOOR CONDITION
Gradient : WITHIN 5 degrees
LOGIQ
α
200 Weight
76 kg (168lbs) without Accessories
POWER SOURCE
230VAC, 50Hz, SINGLE PHASE
For Europe Version
115V, 60Hz, SINGLE PHASE
For USA Version