ELECTRICAL SPECIFICATIONS
Electrical conduit, junction boxes, outlets, circuit breakers, and switches should be in place before installing the
LOGIQ
α
200 console.
3–3–1
Power Supply
Voltage setup is performed in the factory. Different power cables and circuit breakers are used for the 100 (115)
Vac and 220 (240) Vac versions.
3–3–2
Facility Power Receptacle
A separate power outlet with a 10 amp circuit breaker for 100 (115) Vac units, or a 5 amp circuit breaker for 220
(240) Vac units, is recommended. The specific power receptacle used depends on your country’s power line
standards.
The receptacle should have International Electrotechnical Commission (IEC) approval, or equivalent.
3–4
STORAGE AND OPERATION REQUIREMENTS
Probes and peripherals are shipped in separate containers. Shipping weight is approximately 210
lbs (96 kg).
The size of the container is L82 cm x W53 cm x H140 cm (32 in. x 21 in. x 55 in). Table 3–1 provides a summary
of temperature, atmospheric pressure, and humidity tolerances for shipping, installation, and operation.
TABLE 3– 1
STORAGE AND OPERATION REQUIREMENTS
PARAMETER
STORAGE
OPERATION
TEMPERATURE (
C)
(
F)
ATMOSPHERIC PRESSURE
(hPa)
HUMIDITY (%)
(Non–condensing)
–10 to 60
14 to 140
10 to 40
50 to 104
700 to 1060
700 to 1060
5 to 90
5 to 90