6–3–2
Replacement Procedures (Continued)
(A) F1 and F2 on HV Assy
CAUTION
Make sure that only fuse with the required rated current and of the same type is used for
replacement.
1. Turn the system power OFF.and unplug the power cord from the outlet. Follow the OSHA Lockout/Tagout
Requirements (29 CFR 1910.147, Direction 46–015902).
2.
Remove the Left Cover (FRU 404). Refer to 6–2–27.
3.
Remove the Right Cover(FRU 405). Refer to 6–2–28.
4.
Remove the Front Base Cover (FRU 406). Refer to 6–2–29.
5.
Remove the Front Cover (FRU 407). Refer to 6–2–30.
6.
Replace the defective fuse(s) with new one(s). For the location of fuses, refer to ILLUSTRATION 6–55.
7.
Re-install the Covers.
F1 1A 250V
F2 0.5A 250V
HV ASSY
ILLUSTRATION 6–55