PDM module issues
11.8.8 CARESCAPE Network communication issues
Traffic types
Two main types of communication occurs in the CARESCAPE Network: Broadcast and Unicast.
•
Broadcast traffic is sent from one device to all devices on the network. Examples of
CARESCAPE broadcast traffic are device discovery, alarms, and time synchronization.
•
Unicast traffic is sent from one device to another specific device on the network. An
example of CARESCAPE unicast traffic is patient waveforms.
Flow
•
Upstream broadcast: The patient monitor sends broadcasts to other network devices.
•
Downstream broadcast: The patient monitor receives broadcasts from other network
devices.
Possible cause
Recommended action
DC/DC board failure
The VSYS supply voltage for the PDM module is generated in the DC/DC
board. Check from the Webmin
Hardware Statistics
that the supply
voltages are within specification:
- VSYS voltage (mV)
- VSYS_EPORT(mV)
- VSYS_PDM (mV)
If PDM is connected to PDM dock, DC/DC board should identify PDM and
enable the VSYS_PDM circuit breaker on CPU carrier board to connect
VSYS voltage to VSYS_PDM.
If PDM is connected to ePORT, DC/DC board should identify PDM and
enable the VSYS_EPORT circuit breaker on CPU carrier board to connect
VSYS voltage to VSYS_EPORT.
If VSYS voltage is out of spec, the problem is most likely in the DC/DC
board. Replace the DC/DC board.
CPU assembly failure
The Ethernet communication for the PDM module is managed by the CPU
carrier board.
Check PDM Log for possible PDM errors messages, see section
If the cause is none of the above, the problem is most likely in the CPU
carrier boar. Replace the CPU assembly.
Summary of Contents for CARESCAPE Monitor B450
Page 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Page 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Page 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Page 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Page 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Page 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Page 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Page 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Page 175: ...Troubleshooting 167 2062973 003 ...
Page 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Page 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Page 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Page 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Page 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Page 235: ......