Troubleshooting
155
2062973-003
Identical
IP address
noticed
A patient monitor with the
identical IP address is on the
network.
Disconnect the patient monitor that
has the identical IP address.
or
Change the IP address of the patient
monitor that has the duplicate IP
address.
Service Monitor
Error Code
0xHOST1001
One of the internal
temperature sensors indicate
the inside temperature of the
patient monitor is out of
specification. The message
stays on screen as long as the
error condition is valid.
If temperature is too high:
1. Turn off the patient monitor.
2. Let the patient monitor cool down.
3. Check that the ventilation holes are
not obstructed.
4. Ensure that the patient monitor is
installed to a location that meets
the specified environmental
requirements of operating
temperature.
5. Investigate the patient monitor
thoroughly for potential short
circuits and other electrical faults.
6. If possible, log in to Webmin and
select
Diagnostics> Hardware
Stats
to identify the root cause for
the error message.
If temperature is too low (or high):
If the patient monitor has been
transported or stored outside the
operating temperature range, allow it
to stabilize back to operating
temperature range before applying
power.
Service Monitor
Error Code
0xHOST1002
One of the internal supply
voltages is out of the
specification. The message
stays on screen as long as the
condition is valid.
Log in Webmin and select
Diagnostics
> Hardware Stats
to identify the supply
voltage that is below or above the
specification limit.
Service Monitor
Error Code
0xHOST1004
Disk usage exceeds 90%.
1. Back up the clinical and platform
settings and print the licensing
information page via Webmin.
2. Re-install software.
Service Monitor
Error Code
0xHOST1100
The CPU timekeeper battery is
empty. Time and date may be
reset to factory settings.
Readjust time and date first. Replace
the CPU timekeeper battery.
Message
Possible causes
Possible solutions
Summary of Contents for CARESCAPE Monitor B450
Page 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Page 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Page 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Page 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Page 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Page 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Page 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Page 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Page 175: ...Troubleshooting 167 2062973 003 ...
Page 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Page 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Page 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Page 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Page 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Page 235: ......