System overview
23
2062973-003
The patient monitor performs integrity checking on the root file system to detect any
changes to the file system contents. Any modification to the root file system contents will
generate an error to the patient monitoring software application. The patient monitoring
software will then display a technical alarm to the user.
•
Device design and configuration (hardening)
The patient monitor has been hardened through the restriction and removal of user
access to core operating system functionality. In addition, unneeded functionality has
been removed or restricted.
•
Antivirus software
To provide seamless real-time patient monitoring, the patient monitor does not have
antivirus software.
•
Security updates and patching processes
Security updates and patches cannot be applied to the CARESCAPE product without
going through GE’s vigorous software verification and validation process. Any software
update needs will be communicated by GE.
3.5.2 Security operations
Network security
GE requires that the MC port of the patient monitor be connected to a physically or virtually
dedicated CARESCAPE Network MC or S/5 Network, isolated from all other networks.
GE requires that the IX port of the patient monitor be connected to a physically or virtually
dedicated CARESCAPE Network IX with controlled connection to the organization’s general
purpose computing network. Traffic between the organization’s network and IX port of the
patient monitor must be limited to the following packet flows listed below.
Inbound
Packets that are part of the communication initiated by authorized devices in the
organization’s network are allowed to go out of the IX Network (reflexive).
Source device Destination device
Protocol
Destination port Use
Any
Patient monitor
icmp
N/A
ping
Customer
defined
tcp
10000
Webmin
Customer
defined
tcp
10001
Software
transfer
DHCP server
tcp
67, 68
DHCP
Summary of Contents for CARESCAPE Monitor B450
Page 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Page 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Page 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Page 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Page 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Page 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Page 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Page 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Page 175: ...Troubleshooting 167 2062973 003 ...
Page 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Page 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Page 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Page 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Page 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Page 235: ......