Disassembly and reassembly
189
2062973-003
3. Detach the WLAN card from the CPU carrier
board, if installed.
a. Remove the two screws (T6) that hold the
WLAN radio card to the CPU carrier board.
Torque [0.3 Nm].
b. Detach the WLAN radio card carefully
from its connector in the CPU carrier
board.
NOTE: Be careful not to damage the fragile WLAN
antenna boards or antenna cables.
4. Detach the grounding screw (T10) holding the
CPU assembly and AC/DC unit together.
Torque [1.6 Nm].
NOTE: During the reassembly, make sure to
fasten the grounding screw properly. This is
important to ensure that the CPU assembly is
properly connected to the protective earth.
5. Release the latch holding the CPU assembly
and pull the assembly out from the rear unit.
6. Carefully detach the uDOM from the CPU
assembly. The uDOM is connected to the CPU
assembly with a connector from one and with
a nylon support post from the other end. You
may need pliers to squeeze the nylon support
post a little to detach the uDOM.
NOTE: If replacing the CPU assembly, detach
the original uDOM from the CPU carrier board
and then attach it to the new CPU assembly.
The software, licenses and clinical and service
configurations are stored in the original
uDOM.
Summary of Contents for CARESCAPE Monitor B450
Page 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Page 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Page 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Page 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Page 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Page 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Page 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Page 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Page 175: ...Troubleshooting 167 2062973 003 ...
Page 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Page 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Page 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Page 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Page 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Page 235: ......