CARESCAPE Monitor B450
164
2062973-003
11.8.6 Recorder issues
11.8.7 Acquisition module problems
Problem: an acquisition module does not work with the patient monitor.
Locate first whether the problem is in the patient monitor or in the acquisition module:
1.
Connect another, similar, known good module to the suspect patient monitor and check if
the module works normally:
−
“Yes” => The suspect module is most likely faulty. Refer to Module Frames and
Modules Technical Manual for troubleshooting instructions.
−
“No” => The problem is most likely in the patient monitor. Continue troubleshooting
the problem according to the related troubleshooting chart below.
OR
Connect the suspect acquisition module to another, similar, known good patient monitor and
check if the module works normally:
−
“Yes” => The problem is most likely in the patient monitor. Continue troubleshooting
the problem according to the related troubleshooting chart below.
−
“No” => The suspect module is most likely faulty. Refer to Module Frames and
Modules Technical Manual for troubleshooting instructions.
Problem
Possible cause
Recommended action
Recorder does not work.
Graph location is not configured
correctly.
Check the configuration:
Monitor Setup >
Printing > Devices > Setup.
Recorder unit failure.
Replace the recorder unit.
CPU board failure
The serial communication for the recorder
is managed by the EMBC section of the
CPU carrier board.
- Check EMBC Frame logs for possible
EMBC failures.
If the cause is none of the above, the
problem is most likely in the CPU
assembly. Replace the CPU assembly.
Recorder works but nothing
appears on the paper.
Paper installed upsidedown.
Turn the paper roll over.
To test which side is active:
- Place the paper on a hard surface and
draw a line with a fingernail - a dark
line will appear on the active (thermal)
side.
Summary of Contents for CARESCAPE Monitor B450
Page 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Page 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Page 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Page 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Page 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Page 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Page 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Page 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Page 175: ...Troubleshooting 167 2062973 003 ...
Page 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Page 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Page 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Page 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Page 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Page 235: ......