Troubleshooting
145
2062973-003
11.3.2 Ping a TCP/IP network device
Use this Webmin feature to verify connectivity with a network device on the MC Network and IX
Network.
System power (mW)
System power is the total power consumption of the patient
monitor.
Power consumption is measured from the DC/DC
board.
DC/DC temperature (°C)
The internal temperature of the patient monitor
measured by the temperature sensor in the DC/DC board.
CPU Carrier Board
Measurement
Description
+ 12 Vsync CPU Voltage (in mV)
- 12 Vsync CPU Voltage (in mV)
These voltages are generated for the ECG and Invasive
pressure analog outputs and marker-out synchronization
pulse. The voltages are measured by the A/D converter in
the CPU Carrier board.
+15V MOD voltage (mV)
This voltage is generated by the voltage converter in the
DC/DC board. It supplies power to the E-module. The
voltages are measured by the A/D converter in the CPU
carrier board.
VSYS (mV)
This voltage is generated by the voltage converter in the
DC/DC board. It is measured by the A/D converter in the
CPU carrier board.
VSYS_EPORT (mV)
The VSYS_EPORT voltage is supplied for a PDM module
that is connected to the ePort connector DB9) in the rear of
the patient monitor. It is measured by the A/D converter in
the CPU carrier board. The value is shown as “--” if PDM
module is not connected, or if VSYS_PDM voltage is
enabled.
VSYS_PDM (mV)
The VSYS_PDM voltage is supplied for a PDM module that
is connected to the PDM connector in the module rail unit.
It is measured by the A/D converter in the CPU carrier
board. The value is shown as “--” if PDM module is not
connected, or if VSYS_EPORT voltage is enabled.
CPU carrier temperature (°C)
The internal temperature of the patient monitor measured
by the temperature sensor in the CPU carrier board.
CPU Module
Measurement
Description
CPU module temperature (°C)
The internal temperature of the patient monitor measured
by the temperature sensor in the CPU module.
DC/DC Board
Measurement
Description
Summary of Contents for CARESCAPE Monitor B450
Page 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Page 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Page 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Page 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Page 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Page 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Page 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Page 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Page 175: ...Troubleshooting 167 2062973 003 ...
Page 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Page 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Page 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Page 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Page 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Page 235: ......