![GE HEALTHCARE CARESCAPE Monitor B450 Technical Manual Download Page 86](http://html.mh-extra.com/html/ge-healthcare/carescape-monitor-b450/carescape-monitor-b450_technical-manual_571890086.webp)
CARESCAPE Monitor B450
78
2062973-003
4.
Select
Save
.
These settings take effect immediately after they are submitted.
Settings to view 12SL data
1.
Log in to Webmin.
2.
Select
Configuration
>
MUSE/12SL
.
3.
In the
MUSE/12SL
window, enter the following:
4.
Select
Save
.
These settings take effect immediately after they are submitted.
7.8 Admit settings
7.8.1 Patient ID prefix
The patient monitor will automatically generate a temporary, unique patient ID when a patient
with unknown ID is admitted to the patient monitor. The patient monitor will use this temporary
patient ID for all 12SL reports that are sent to MUSE until the patient is discharged from the
patient monitor, or his/her patient ID is changed. The temporary patient ID is generated from
the temporary patient ID prefix, care unit name, bed name, and current time.
The temporary patient ID prefix is a hospital defined prefix that is used as the first two
characters in a temporary patient ID to ensure its uniqueness inside the hospital.
1.
Log in to Webmin.
2.
Select
Configuration
>
Admit Settings
.
3.
In the
Sub-Modules for Admit Settings
menu, select
Patient ID Prefix
.
4.
In the
Patient ID Prefix
window, enter a 2-character prefix.
NOTE: Valid values are uppercase letters and numbers.
5.
Select
Save
.
All changes take effect immediately.
- Location ID
Identifies the location ID number (within the range 0 to 599)
associated with the patient monitor for searching the MUSE
system.
- Site Number
Identifies the site number (within the range 1 to 254) associated
with the patient monitor for searching the MUSE system.
- MUSE Web Username
Username used to authenticate with the MUSE
Web to access 12-lead reports.
- MUSE Web Password
Password used to authenticate with the MUSE
Web to access 12-lead reports.
- Confirm Password
The password and the confirmed password must
be the same.
- MUSE Web URL
Used to locate the MUSE Web system to access
12-lead reports. Enter the URL in a valid format.
Summary of Contents for CARESCAPE Monitor B450
Page 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Page 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Page 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Page 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Page 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Page 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Page 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Page 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Page 175: ...Troubleshooting 167 2062973 003 ...
Page 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Page 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Page 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Page 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Page 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Page 235: ......