Troubleshooting
151
2062973-003
3.
Select the log file you want to view.
4.
Select the information you want to view.
•
For the Webmin Action Log, select the user, module, and timeframe and select
Search
.
•
For the other types of logs, select the link associated with the information you want
to view.
11.4 Power management LEDs
The DC/DC board includes troubleshooting LEDs that helps the user to troubleshoot start-up
related problems in the patient monitor.
The troubleshooting LEDs in the DC/DC board are visible through the ventilation holes in the
rear of the patient monitor (see picture below).
LED
Description
On/standby (red, H6)
The red on/standby LED should be lit momentarily when the on/
standby button is pressed. A lit LED normally indicates that the DC/DC
power supply board recognizes that the user has pressed the on/
standby button and should turn the patient monitor On/Off.
If the On/standby LED is not lit when the on/standby button is
pressed, check the patient monitor for any of the following problems:
•
Faulty user interface board.
•
User interface cable is not connected to DC/DC board.
•
Faulty DC/DC board.
Summary of Contents for CARESCAPE Monitor B450
Page 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Page 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Page 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Page 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Page 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Page 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Page 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Page 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Page 175: ...Troubleshooting 167 2062973 003 ...
Page 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Page 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Page 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Page 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Page 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Page 235: ......