Disassembly and reassembly
197
2062973-003
•
Mid-frame assembly
Reassemble in reverse order:
- Ensure that the display gasket is correctly aligned and that the four snaps lock properly when attaching the
front unit assembly back to the mid-frame assembly.
- Guide the touchscreen and user interface cables through the openings in the mid-frame assembly.
Connect the cables to the connectors in the DC/DC board.
NOTE: You may find it easier to first connect the user interface cable to the DC/DC board and guide it towards
the user interface board through the opening in the mid-frame assembly.