A - 2 (4)
2062973-003
b.) Impedance of protective earth
connection
without power cord
≤
0.1 ohms
with power cord
≤
0.2 ohms
Normal Condition (NC)
≤
500 µA
EN /IEC
≤
300 µA
UL
Single Fault Condition (SFC)
≤
1 mA
Enclosure leakage current (touch
current) test
Normal Condition (NC)
≤
100 µA
Single Fault Condition (SFC)
≤
500 µA
EN /IEC
≤
300 µA
UL
Module type
S/N
Observed result
Acceptance
criteria
PASS
N.A.
FAIL
ECG
SpO2/
InvP
Patient (source) leakage current tests, using a test body
Normal Condition (NC)
Polarity: NORMAL
≤
10 µA
Single Fault Condition (SFC)
Ground closed (normal)
≤
50 µA
Ground open (normal)
≤
50 µA
Ground closed (reverse)
≤
50 µA
Ground open (reverse)
≤
50 µA
Normal Condition (NC)
Polarity: Reserved
≤
10 µA
Patient (sink) leakage current tests
Polarity: NORMAL
≤
50
µA
Polarity: Reserved
≤
50
µA
Module type
S/N
Observed result
Acceptance
criteria
PASS
N.A.
FAIL
Patient (source) leakage current tests, using a test body
Normal Condition (NC)
Polarity: NORMAL
≤
10 µA
Single Fault Condition (SFC)
Ground closed (normal)
≤
50 µA
Ground open (normal)
≤
50 µA
Ground closed (reverse)
≤
50 µA
Ground open (reverse)
≤
50 µA
Normal Condition (NC)
Polarity: Reserved
≤
10 µA
Patient (sink) leakage current tests
Polarity: NORMAL
≤
50
µA
Polarity: Reserved
≤
50
µA
Summary of Contents for CARESCAPE Monitor B450
Page 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Page 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Page 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Page 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Page 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Page 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Page 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Page 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Page 175: ...Troubleshooting 167 2062973 003 ...
Page 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Page 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Page 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Page 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Page 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Page 235: ......