6-2
897N203029A FDC NX600 Instruction Manual 2022.01
6 Maintenance
6.1 Periodic Maintenance Items
6.1.1 Items for Daily and Periodic Inspection by the User
Users must perform the daily and periodic maintenance and replacement of parts in the following table.
Location
Cleaning interval
Inspection
interval
Replacement interval
Procedure
Air filter
Once a month
-
-
Refer to
Transfer bar
Once every 3 months,
or as appropriate, such
as when inaccurate
measurement values
occur
-
-
Refer to
Incubator
Once every 3 months,
or as appropriate, such
as when inaccurate
measurement values
occur
-
-
Refer to
Sample application unit
As appropriate, such as
when a slide transfer
error or tip ejection error
occurs
-
-
Refer to
ISE unit
Once every 3 months,
or as appropriate, such
as when inaccurate
measurement values
occur
-
-
Refer to
Recording paper
-
-
Appearance of red lines on both
sides of paper
Refer to
Light source lamp
-
-
When a lamp replacement error
occurs, or when the cumulative
lighting time exceeds 1,000 hours
Refer to
Sampler O-ring
Once a week
Once a month
Once a year
Refer to
Slide reading section
When slide reading
errors occur frequently
-
-
Refer to
Reference fluid cap
packing
When replacing refer-
ence fluid
-
Once every 3 months
Refer to
PF sensor
As appropriate, such as
when errors related to
PF extraction occur
-
-
Refer to
PF reference plate
-
-
PF Suction Pad
Once a month
Once every 2 years
PF Tube
-
-
Once every 2 years or when
the power has been off for one
month
Refer to
WARNING:
After inspection, always connect the power cord, switch on the power, and then check operation.
6