9-25
897N203029A FDC NX600 Instruction Manual 2022.01
9 Function Menu
9.3.10 Lot Compensation Coefficients
Enter the lot compensation coefficients (c, d, and e) printed on the QC cards of slides. This mode is
needed if it is not possible to read the data directly because the QC card was lost or spoiled.
IMPORTANT:
Incorrect inputs for (c, d, and e) will cause incorrect measurement results.
Make sure that the inputs for (c, d, and e) are correct using this mode.
IMPORTANT:
When the slide type number was changed, even if the new values (c, d, and e) are entered by us-
ing this setting, the new standard curve information will not be updated. It is necessary to read the
QC card when the slide type number was changed.
The slide type number is the most significant digit of a lot number.
(Example) The slide type number of “Lot No.123456” is 1.
Tap the button of the “Measurement item - sample
type” you want to edit.
NOTE
Only “Measurement item - sample type”
items that have been read from QC cards
are displayed as buttons.
NOTE
Immuno-chemical items are not displayed.
NOTE
Measurement items are displayed in as-
cending order of test code, and sample type
is displayed in the following order: [PS] and
[W].
9