6-4
897N203029A FDC NX600 Instruction Manual 2022.01
6 Maintenance
6.2 Cleaning the Air Filters
The air filters should be checked and cleaned once a month.
IMPORTANT:
If the air filters are not cleaned, an increase in the temperature inside the analyzer may negatively
affect the measurement values.
NOTE
Clean all 4 filters.
1. Switch off the analyzer power.
Louver
Louver
1
2
2. Remove the four louvers and pull out the
air filters located inside of the louvers.
Lift the louver upward
1
and then pull forward
2
to remove.