6 Maintenance
6.1.2 Items for Maintenance and Inspection by Repair Personnel
Designated repair personnel must perform the maintenance and replacement of parts in the follow-
ing table.
Parts
Replacement interval
Interference filter
Once every 4 years
Syringe
Once every 3 years
Incubator pressure plate
Inspection once every 3 years
WARNING:
After inspection, always connect the power cord, switch on the power, and then check operation.
NOTE
The standard number of slides used per year for this analyzer is assumed to be 7,200 col-
orimetric slides and 240 ISE slides. (In the case of 240 days of use per year, 30 colorimetric
slides/day and one ISE slide/day)