8 Troubleshooting
8-42
897N203029A FDC NX600 Instruction Manual 2022.01
8.2.15 Other Errors
Error
code
Error message
Error condition
Troubleshooting
W0020
No disposal box
Put Disposal box to
the end and then press
START key.
Measurement was
started with the dis-
posal box pulled out.
Insert the disposal box completely, and
then start measurement.
E0021
Disposal box open
ERR
Disposal box opened
during measurement.
Close Disposal box
and run the tests from
the beginning.
The disposal box was
open during measure-
ment.
Close the disposal box, and then start
measurement again from the begin-
ning.
W0022
Disposal box full
Disposal box became
full of slides and tips.
Discard slides and tips
to empty the box when
measurement is not in
progress.
The disposal box is full
of slides and tips.
While a measurement is not being per-
formed, discard the slides and tips in
the disposal box.
NOTE
When there are 150 slides or
75 tips in the disposal box, this
error occurs. If the disposal box
is pulled out, it is determined
that the slides and tips were
discarded, and the disposal
count is reset. Always empty
the disposal box after pulling it
out.
E0080
Sample tip sensor
ERR
Contact your dealer.
Malfunction of the
sample tip sensor.
Please contact your dealer for assis-
tance.
W0500
Rear fan 1 stop
Contact your dealer.
A fan error has oc-
curred.
W0501
Rear fan 2 stop
Contact your dealer.
W0502
Intake fan stop
Contact your dealer.
W0503
Exhaust fan stop
Contact your dealer.
I5920
Turn the power ON
30 days or more have
passed from when the
analyzer was started.
For stable system
performance, turn the
power ON.
A long time has passed
since start-up.
Switch the analyzer off and on.
8