9-36
897N203029A FDC NX600 Instruction Manual 2022.01
9 Function Menu
9.3.17 Control Fluid Setting
Configure the control liquid name / lot number and acceptable range.
1. Tap the button of the control liquid to be
edited.
NOTE
[-] (control liquid not registered) of Step
1 cannot be edited.
2. Tap a [Lot] button to choose a lot.
3. If you touch a display section (lower limit
or upper limit) of the acceptable range to
be edited, a software keyboard (Refer to
Section 2.3) appears so you can edit the
value.
IMPORTANT:
If the entry of the acceptable range is incorrect,
the judgement of the control measurement results
is incorrect. Always make sure that the entered
values are correct.
NOTE
The units configured by the [Unit conver
-
sion] of the function menu are displayed.
NOTE
If the acceptable range is changed, in-
formation from before the change is not
saved.
4. To edit a lot number, touch the [Lot edit-
ing] button, and then enter the number
using the software keyboard that ap-
pears.
NOTE
To edit or add a control liquid name,
tap the button of a control liquid name,
and then enter the information using the
software keyboard that appears.
9