8 Troubleshooting
8-23
897N203029A FDC NX600 Instruction Manual 2022.01
Error
code
Error message
Error condition
Troubleshooting
E1502
Temperature control
ERR
Incubator temperature
sensor disconnected.
Check the incubator
cable. Contact your
dealer.
When the measure-
ment results are ap-
pended with a “$” mark
or an error listed to the
left occurs, the incuba-
tor cable either is not
connected or has been
severed.
Switch off the power and connect the
incubator cable again.
NOTE
After inserting the connector of
the incubator cable, you must
turn it to the right to lock it.
NOTE
If the error occurs again, please
contact your dealer for assis-
tance.
E1503
Temperature control
ERR
Incubator temperature
sensor damaged.
Contact your dealer.
E1504
Temperature control
ERR
Incubator heater dis-
connected.
Check the incubator
cable. Contact your
dealer.
E1512
ISE temperature con-
trol ERR
ISE temperature con-
trol element damaged.
Contact your dealer.
Disconnection error of
the ISE incubator
Please contact your dealer for assis-
tance.
E1514
ISE temperature con-
trol ERR
ISE heater disconnect-
ed.
Contact your dealer.
8