7 Quality Control
7-3
897N203029A FDC NX600 Instruction Manual 2022.01
5. Tap the button of the measurement item
to be viewed.
Acceptable
range
lower limit
Show/hide
button
Measurement
result average
Acceptable
range
upper limit
Outside of display
area
Control lot was
changed
Control lot was
changed
Outside of accept-
able range
6. The control chart appears.
NOTE
If the control liquid button is tapped, you
can change the control liquid to be dis-
played.
NOTE
Measurement results of outside of
measurement range (@), disposal box
open (*), and expired slide (#) are not
reflected in the control chart.
NOTE
If the measurement date is touched and
selected, and the show/hide button is
touched, the plot of these measurement
results can be hidden (or shown).
7