4-24
897N203029A FDC NX600 Instruction Manual 2022.01
4 Applicable Usage Methods
4.4 Retesting
You can test again using the previous sample number, sample ID, sample type setting, and refer-
ence interval setting by tapping the (retest) button at the measurement results screen. Further-
more, when a measurement cannot be performed due to an error, you can skip the re-entry and re-
configuration of the setting information by tapping the
(retest) button. When a PF measurement
is run again, the analyzer does not filter using the PF, but instead uses plasma that was previously
filtered by the PF.
IMPORTANT:
The (retest) button is designed only to automatically enter the sample number and sample ID of
the previously measured sample. Do not use this button to retest samples from before the previous
sample.
CAUTION:
Do not tap the [START] button when the sample has already been filtered by the PF. If the [START]
button is tapped, the analyzer may malfunction.
Retesting after filtering by the PF
• When you will rerun a measurement using the PF, do not use the [START] button to restart the
measurement. Always use the (retest) button.
•
Make sure that the sample volume is sufficient.
•
Make sure that the liquid surface of the filtered sample is level.
IMPORTANT:
Do not retest GPT (ALT) of PF filtered plasma because its activity decreases over time.
4