6-14
897N203029A FDC NX600 Instruction Manual 2022.01
6 Maintenance
Tapered part
8. Wipe the tapered part of each pres-
sure plate (the metallic luster taper part
around the black part).
NOTE
Wipe the entire circumference as blood
stains may be difficult to see by eye.
Surface
9. Wipe the surface (black part) of each
pressure plate.
NOTE
Wipe gently to avoid scratching, as it
will affect the measurement values.
IMPORTANT:
When wiping, change the cloth surface for each
cell. Otherwise, dirt may be transferred.
IMPORTANT:
Do not use a cotton swab.
The components of the cotton swab may nega-
tively affect the measurement values.
IMPORTANT:
After cleaning, do not touch the pressure plate sur-
face with your fingers.
This may affect the measurement values.
Measurement cell
10. Wipe the lower part (measurement cells)
in the same way.
6