6-40
897N203029A FDC NX600 Instruction Manual 2022.01
6 Maintenance
RE
5. Turn the PF Suction Pad in the direction
shown and remove it while loosening.
6. Install the new PF Suction Pad by turn-
ing it in the opposite direction.
NOTE
Be sure to tighten the PF Suction Pad
securely.
7. Switch on the analyzer.
8. Run the leak check.
If the check results are [NG], perform the re-
placement again.
6