897N203029A FDC NX600 Instruction Manual 2022.01
4 Applicable Usage Methods
4.3.2 Blood Collection Tubes for Tests Using a PF
NOTE
The FDC NX600i cannot measure using a PF.
The following blood collection tubes can be used with PF measurements: φ16×100 mm, φ13×100
mm, and φ13×75 mm.
(Outer diameter of Ø16 indicates Ø15.0 to 16.6 mm and Ø13 indicates Ø12.0 to 13.3 mm.)
IMPORTANT:
Use blood collection tubes containing Heparin.
IMPORTANT:
Refer to the “Instructions for Use” of the plasma filter (PF) and always follow the instructions.
NOTE
Do not use blood collection tubes that may cause the tip of the sample suction nozzle of the
PF to be physically blocked by additives or other materials.
NOTE
As shown in the illustration below, the required sample volume is 6.5 mL or more for
Ø16×100 mm tubes, 6 mL or more for Ø13×100 mm tubes, and 3 mL or more for Ø13×75
mm tubes.
0023
PF
10mmの空き
スペースが必要です
10mmの空き
スペースが必要です
測定可能範囲
測定可能範囲
検体は
φ16x100mm管で6.5ml以上
φ13x100mm管で6ml以上
必要です
検体は
φ13x75mm管で3ml以上
必要です
100mm
75mm
10 mm of empty
space is necessary
10 mm of empty
space is necessary
The required sample volume
is 6.5 mL or more for Ø16×100
mm tubes, and 6 mL or more
for Ø13×100 mm tubes.
The required sample
volume is 3 mL or more for
Ø13×75 mm tubes.
Measurable
PF
100 mm
75 mm
Measurable
NOTE
Use PF sample racks with the following labels for tests using a PF.
4