6-37
897N203029A FDC NX600 Instruction Manual 2022.01
6 Maintenance
3. Follow the on-screen instructions to load
the PF sample rack and PF leak check
tool to the sample position, and then tap
the [START] button to run the leak check.
If check ends, the Check results screen appears.
CAUTION:
Keep your fingers clear as the PF part moves.
4. Check the results of the [Leak check].
If the check results are [OK], it is not necessary
to clean or replace the pad.
NOTE
The check results can be printed by tap-
ping the
button.
If the check results are [NG], tap the [Cleaning]
button, clean the PF unit, and then run the leak
test again.
If this does not resolve the problem, tap the
[Procedure] button, and then check the assem-
blies of the PF pump tube or replace the tube,
referring to the displayed instructions.
NOTE
If the [Result details] button is tapped,
the Advanced leak check results display
screen appears.
NOTE
The check results can be printed by tap-
ping the
button.
5. Once the check ends, remove the PF
sample rack and PF leak check tool.
6