6 Maintenance
6.3 Inspecting and Cleaning the Transfer Bar, Incubator, Sam-
ple Application Unit, and ISE Unit
Measuring results are affected by stains (from blood samples, etc.) inside of the incubator and ISE
unit. These parts should be checked and cleaned at least once every three months. When a slide
transfer error or a tip ejecting error occurred, clean the sample application unit.
WARNING:
Always wear gloves, a lab coat, and safety glasses while working.
If you accidentally touch a contaminant, immediately rinse the contaminated area thoroughly un-
der running water, and then disinfect the area.
Seek medical assistance if necessary.
WARNING:
Dispose of contaminants used for cleaning as infectious waste.
IMPORTANT:
Staining of the incubator, spotting parts, and electrolyte parts with samples such as blood may af-
fect the measured values.