6 Maintenance
6.7 Cleaning the Slide Reading Section
When a slide-detection error has occurred, dust and other soiling may be the cause. Clean the slide
reading section to avoid these errors.
WARNING:
Always wear gloves, a lab coat, and safety glasses while working.
If you accidentally touch a contaminant, immediately rinse the contaminated area thoroughly un-
der running water, and then disinfect the area.
Seek medical assistance if necessary.
WARNING:
Dispose of contaminants used for cleaning as infectious waste.
RE
1. Switch off the analyzer power.
2. Open the sampler cover and panel unit.
3. Remove the slide cartridge.
RE
4. Using your finger, move the transfer bar
under the right cover.
RE
5. Air blow from the gap where the slide
cartridge is placed, as shown in the left
figure.
6