69
Note: Digital oscilloscopes are susceptible to aliasing which can cause inaccurate signal reconstruction.
The Maximum Recommended Input frequency for each horizontal scale setting is shown in the display
area. For best results follow the guidelines shown in the table below:
Horizontal
Scale
Maximum
Signal
Frequency
Maximum
Recommended
Signal Frequency
20 us
50000 Hz 50000
Hz
50 us
50000 Hz 20000
Hz
100 us
50000 Hz 10000
Hz
200 us
25000 Hz 5000
Hz
500 us
10000 Hz 2000
Hz
1 ms
5000
Hz 1000
Hz
2 ms
2500
Hz 500
Hz
5 ms
1000
Hz 200
Hz
10 ms
500
Hz 100
Hz
20 ms
250
Hz 50
Hz
50 ms
100
Hz 20
Hz
100 ms
50
Hz 10
Hz
200 ms
25
Hz 5
Hz
500 ms
10
Hz 2
Hz
1 s
5
Hz 1
Hz
Marker Mode
Provides display marker control with a horizontal soft key menu. Markers can be turned off or on with a
choice of numeric readout for the signal measurements as follows:
Delta V – The numeric reading shows the difference in amplitude between marker positions.
Delta T – The numeric reading shows the difference in time between marker positions.
1/DeltaT – The numeric reading shows the inverse of the time difference between markers which can be
used to determine the frequency of a repetitive waveform.
Absolute – The numeric reading shows the absolute amplitude at the marker position
Toggle Marker
Cycles through the available markers to select the active one (yellow) for positional adjustment on the
display. The active marker is moved with the Left/Right (◄►) keys
Trigger Mode
Submenu selections for the horizontal sweep trigger mode. The choices are as follows:
Summary of Contents for R8100 SERIES
Page 19: ...19 Figure 2 1 2 Left and Right Sides of the R8100...
Page 164: ...164 Figure 2 2 7 7 5 Power Profile Frame with Markers...
Page 224: ...224 Figure 3 2 9 6 1 Submenu after pressing Options soft key in the Settings menu...
Page 243: ...243 PTC Test in Monitor Mode Eye Diagram Display...
Page 244: ...244 Power Profile Display Linear and Circular Constellation Displays...
Page 245: ...245 Symbol Distribution Plot...