Require Operator ID (Test Preferences submenu)
AutoTune requires the Operator ID to be set before testing can be initiated because it is logged in the test
data and reported in the formatted test report. If the field is blank, the operator is not recorded. Setting this
preference to No will allow testing to be started without an Operator ID.
Recommended: Site preference
Default: Yes
Display Report (Test Preferences submenu)
AutoTune stores the results from each test in a comma delimited log file. These results can be viewed at
any time by clicking on the Tools’ “Test Reports” button on the AutoTune screen. The button displays the
logged test results in an easy to read format. AutoTune will automatically display this screen at the
completion of all tests.
Setting this preference to No will prevent the Test Report from being displayed when testing is complete.
Recommended: Site preference
Default: Yes
Unique Log Files (Test Preferences submenu)
Test log data is recorded and saved to a comma-delimited format filename based on the serial number of
the radio under test (e.g. XTS5000_11124.csv). Setting this preference to No will cause all test log data
to be saved under a common log file name based on the radio make and model (e.g. XTS5000.csv).
Recommended: Set to No if testing and reporting a batch of radios.
Default: Yes
Display Setup Diagrams (Test Preferences submenu)
This setting controls the display of Test Setup Diagrams at the start of testing. When No, the diagrams
will not be shown.
Note: If the radio under test requires the test setup to be changed during testing, this option will be ignored.
In this situation, it is critical that the user review the diagrams to ensure that the cable setup is correct.
Incorrect setup will result in false test failures, and possibly damage to the radio or analyzer.
Recommended: Yes
Default: Yes
RF Level Offset (Test Preferences submenu)
When enabled, the R8100 applies the RF offset in the RF In/Out field next to it to the output levels and
input power measurements used by AutoTune. When disabled (default), the value in the RF In/Out field
is ignored.
This field mirrors the RF Level Offset field described in System Settings.
Recommended: Site preference
Summary of Contents for R8100 SERIES
Page 19: ...19 Figure 2 1 2 Left and Right Sides of the R8100...
Page 164: ...164 Figure 2 2 7 7 5 Power Profile Frame with Markers...
Page 224: ...224 Figure 3 2 9 6 1 Submenu after pressing Options soft key in the Settings menu...
Page 243: ...243 PTC Test in Monitor Mode Eye Diagram Display...
Page 244: ...244 Power Profile Display Linear and Circular Constellation Displays...
Page 245: ...245 Symbol Distribution Plot...