240
Connect the 50 Ohm load to the open port on the Power splitter and start a calibration sweep by pressing
“Analyze” then “Calibrate”. The calibration sweep takes approximately 1 minute to complete.
Note:
Whenever the Center Frequency, Maximum Length, or Velocity Factor are changed the Cable Fault
instrument will force a recalibration before performing an analysis sweep.
Remove the 50 Ohm load and connect the cable to be tested - choose either Single Sweep or Continuous
in the “Analyze” submenu. Each sweep takes approximately 1 minute to complete and the display will
update with a plot of return loss vs. distance.
Summary of Contents for R8100 SERIES
Page 19: ...19 Figure 2 1 2 Left and Right Sides of the R8100...
Page 164: ...164 Figure 2 2 7 7 5 Power Profile Frame with Markers...
Page 224: ...224 Figure 3 2 9 6 1 Submenu after pressing Options soft key in the Settings menu...
Page 243: ...243 PTC Test in Monitor Mode Eye Diagram Display...
Page 244: ...244 Power Profile Display Linear and Circular Constellation Displays...
Page 245: ...245 Symbol Distribution Plot...