RF Gen Out
Specifies the gains or losses between the RF Gen Out port and the unit under test.
In the case of a loss, such as a cable or attenuator, enter a
negative
value. The affected field values will be
increased
to compensate. For example, if a -6.0 dB value is entered, the Output Level amplitude is
increased
by 6 dB.
In the case of a gain, such as an amplifier, enter a
positive
value. The affected fields will be
decreased
to
compensate. For example, if a 10.0 dB value is entered, the Output Level amplitude will
decrease
by 10
dB.
This value is not used if RF Level Offset is disabled (Off).
Antenna
Specifies the gains or losses between the Antenna port and the unit under test.
In the case of a loss, such as a cable or attenuator, enter a
negative
value. The affected field values will be
increased
to compensate. For example, if a -6.0 dB value is entered, the Spectrum Analyzer trace is
increased
by 6 dB.
In the case of a gain, such as an amplifier, enter a
positive
value. The affected fields will be
decreased
to
compensate. For example, if a 10.0 dB value is entered, the Spectrum Analyzer trace will be
reduced
by
10 dB.
This value is not used if RF Level Offset is disabled (Off).
Reset to Defaults
All R8100 system settings are restored to the factory defaults.
Time
Sets the R8100 system clock in 24-hour notation. Use the numeric keypad and the Up/Down (▲▼), and
Left/Right (◄►) keys to make or change entries.
Date
Sets R8100 system date in month, day, year (MM.DD.YYYY) notation. Use the numeric keypad and the
Up/Down (▲▼), and Left/Right (◄►) keys to make or change entries.
Apply Date/Time Changes
Applies the time and date changes selected above to the R8100 system clock.
2.2.8.3
Network Settings for Remote Operation
The R8100 can be connected to a TCP/IP network for remote operation (see section 3 REMOTE
OPERATION) using the unit’s Remote Front Panel option (R8-Remote). Communication over the
Summary of Contents for R8100 SERIES
Page 19: ...19 Figure 2 1 2 Left and Right Sides of the R8100...
Page 164: ...164 Figure 2 2 7 7 5 Power Profile Frame with Markers...
Page 224: ...224 Figure 3 2 9 6 1 Submenu after pressing Options soft key in the Settings menu...
Page 243: ...243 PTC Test in Monitor Mode Eye Diagram Display...
Page 244: ...244 Power Profile Display Linear and Circular Constellation Displays...
Page 245: ...245 Symbol Distribution Plot...