Figure 2.2.7.3.1-2
Main screen in Monitor mode after pressing the PROJECT 25 soft key
2.2.7.3.1.1
RF Zone
Input Level (RF Zone display)
Input Lvl displays the average power in the P25 channel under test.
Note: When the RF input power on the RF In/Out port is above +20 dBm (100 mW), the R8100 utilizes a
broadband power detector for the measurement. The “Input Lvl” field in the RF Zone changes to “Watt
Meter” to indicate this measurement mode. For best accuracy disable the Pre-amplifier in Monitor Mode,
and set the Gen Port in Generate Mode to RF In/Out.
Freq Error (RF Zone display)
Freq Error displays the frequency difference of the received P25 transmission carrier minus the R8100
Monitor Frequency.