Getting Started Testing Fiber
Tutorial: Running a Loss/Length Test
2
2-19
Step 3: Configuring the Loss/Length Test
Do the following on the tester you will use to initiate
tests (the main tester) and on the remote tester where
indicated:
3-1
From the
SETUP
screen, press
E
Tab
to select the
Loss/Length
tab.
3-2
Press
t
to see the available test limits.
3-3
Use
LM
to highlight
Document only
; then press
t
to select the limit.
3-4
Use
LM
to highlight
REMOTE END SETUP
; then
press
t
.
3-5
Use
LM
to highlight
Smart Remote
; then press
t
.
If a message about setting the reference appears,
press
B
No
. You will set the reference in Step 4.
3-6
Use
LM
to highlight
THIS UNIT
; then press
t
.
3-7
Use
LM
to highlight
Main
; then press
t
.
3-8
Use
LM
to highlight
BI-DIRECTIONAL
; then press
t
.
3-9
Use
LM
to highlight
Disabled
; then press
t
.
3-10
On the remote unit’s
Loss/Length
tab set
REMOTE
END SETUP
to
Smart Remote
and
THIS UNIT
to
Remote
.
Step 4: Setting the Reference
4-1
Clean the tester's and source's OUTPUT connectors
and the reference test cord connectors.
4-2
Press
F
. Use
LM
to highlight
Set Loss/Length
Reference
; then press
t
.
4-3
Connect the main and remote testers as shown on
the
SET REFERENCE
screen and in Figure 2-3; then
press
t
.
4-4
On the
VIEW REFERENCE
screen press
C
OK
. (You
do not need to enter reference test cord lengths for
this tutorial.)
Summary of Contents for OF-500 OptiFiber
Page 12: ...OF 500 OptiFiber Technical Reference Handbook x ...
Page 18: ...OF 500 OptiFiber Technical Reference Handbook xvi ...
Page 27: ...Getting Acquainted Powering the Tester 1 1 9 ajt20f eps Figure 1 1 Battery Pack Features ...
Page 29: ...Getting Acquainted Verifying Operation 1 1 11 ajt56f eps Figure 1 2 Removing the Module ...
Page 46: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 1 28 ...
Page 136: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 4 6 ...
Page 192: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 6 48 ...
Page 254: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 11 36 ...
Page 256: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Page 272: ...B 14 OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Page 274: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook C 2 ...
Page 282: ...OF 500 OptiFiber Technical Reference Handbook 8 ...