OF-500 OptiFiber Certifying OTDR
Technical Reference Handbook
8-6
If the Power Reading is Low
•
Verify that the reference test cord is connected to the
tester's INPUT port and that all other connections are
secure.
•
Verify that the wavelength you selected on the tester
matches the source's wavelength.
•
If you are measuring the output of a source used for
testing, verify that the source's output is set to
continuous wave. The modulated output produces a
lower power reading than the continuous wave
output.
•
Verify that all patch cords and adapters are good and
are of the appropriate type and core size for the
source.
•
If you are measuring the power received through
cabling, use the OTDR to check the cabling for faults.
W
Caution
Disconnect the source before connecting the
OTDR to the cabling. Live sources can damage
the OTDR’s receiver.
•
Clean all fiber connections, then retest.
•
Use a fiber microscope, such as the Fluke Networks
FiberInspector probe, to inspect the fiber connectors.
Polish the connectors if necessary, then retest.
Summary of Contents for OF-500 OptiFiber
Page 12: ...OF 500 OptiFiber Technical Reference Handbook x ...
Page 18: ...OF 500 OptiFiber Technical Reference Handbook xvi ...
Page 27: ...Getting Acquainted Powering the Tester 1 1 9 ajt20f eps Figure 1 1 Battery Pack Features ...
Page 29: ...Getting Acquainted Verifying Operation 1 1 11 ajt56f eps Figure 1 2 Removing the Module ...
Page 46: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 1 28 ...
Page 136: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 4 6 ...
Page 192: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 6 48 ...
Page 254: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 11 36 ...
Page 256: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Page 272: ...B 14 OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Page 274: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook C 2 ...
Page 282: ...OF 500 OptiFiber Technical Reference Handbook 8 ...