Using the Loss/Length Option
Testing in Smart Remote Mode
6
6-23
Table 6-5. Smart Remote Mode RESULTS Screen Features
Item Description
END 1-2
END 2-1
Output
Fiber
Input Fiber
These refer to the direction of the test results. If
THIS END
on the main tester is set to
END 1
,
END 1-2
shows
results for the main tester's output fiber;
END 2-1
shows results for the input fiber. The results are reversed if
the main tester is set to
END 2
. Press
A
to switch between the results for each fiber.
When bi-directional testing is enabled,
Output Fiber
and
Input Fiber
refer to the fiber connections at the
main tester after you swapped the fibers. The tester shows results for both test directions. Press
A
to switch
between the results for each fiber.
PASS:
The measurement in the column is within the limit.
FAIL:
The measurement in the column exceeds the limit.
RESULT
The loss, length, or propagation delay result for the fiber. Propagation delay is half of the time taken for the
signal to travel from the tester's OUTPUT connector to the INPUT connector (minus the delay of the reference
test cords and adapters). "Too High" is shown if the loss is too high to measure.
If loss is negative, set the reference and test the cabling again. See Table 6-10 for details.
LIMIT
The maximum loss, length, or propagation delay allowed by the selected test limit.
MARGIN
The difference between the limit and the measured value. Margin is negative for measurements that fail.
N/A
The results show
N/A
for values not defined by the selected test limit.
Summary of Contents for OF-500 OptiFiber
Page 12: ...OF 500 OptiFiber Technical Reference Handbook x ...
Page 18: ...OF 500 OptiFiber Technical Reference Handbook xvi ...
Page 27: ...Getting Acquainted Powering the Tester 1 1 9 ajt20f eps Figure 1 1 Battery Pack Features ...
Page 29: ...Getting Acquainted Verifying Operation 1 1 11 ajt56f eps Figure 1 2 Removing the Module ...
Page 46: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 1 28 ...
Page 136: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 4 6 ...
Page 192: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 6 48 ...
Page 254: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 11 36 ...
Page 256: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Page 272: ...B 14 OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Page 274: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook C 2 ...
Page 282: ...OF 500 OptiFiber Technical Reference Handbook 8 ...