X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
6
Software Menus
6.2.3 Setup Menu
6.2.3.3.4.1.1
Setup Cross Interference Linearization Curve
Setup..
Measurement..
Cross Interference..
IF Source
1 ... 4
Linearization Curve..
Reference value to normalize the linearization
curve.
Accepted range:
-1E+9 ... +1E+9
Enter up to four polynomial factors to linearize
the interfering component´s input signal with
a fourth-order polynomial of the form
Ch1
IF Source1
Reference Value:
1
IF Polynomial a1:
1.00000000
IF Polynomial a2:
0.00000000
IF Polynomial a3:
0.00000000
IF Polynomial a4:
0.00000000
Note!
The menu figure to the left shows the default
setup for the polynomials, which relates to a
straight line.
So, if your IF source signal is linear, no fur-
ther actions or changes in this menu are
required.