Emerson Process Management GmbH & Co. OHG
4-9
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
4
Installation
Plugs and sockets required for the electrical
connections are on the rear panel of the units
(Fig. 4-4 & 4-5).
X-STREAM XEGP analyzers provide an in-
ternal wide range power supply for worldwide
use.
X-STREAM XEGK analyzers are powered
by an internal wide range power supply for
worldwide use, or by an external DC 24 V
power supply unit, optionally supplied with
4.6.1 X-STREAM XEGK, X-STREAM XEGP
4.6.1 Installation - X-STREAM XEGK, X-STREAM XEGP
Fig. 4-4:
X-STREAM XEGK - Rack Mount Version Rear Panel
Gas inlets
Gas outlets
AC Power inlet
with fuses
Analog outputs / relay
outputs
USB connectors
RJ45 ethernet connectors
Serial interface
( RS 232 /
Modbus 485)
Digital in- /
outputs
Serial CAN
interface
Service interface
the unit. If an external PSU is not included
in delivery, another unit can be used instead,
provided it conforms to the specifications on
X-STREAM XEGK / XEGP analyzers should
be operated in a horizontal position.
Six screws at the front panel enable to install
XEGK models into a rack. The external PSU
is optionally available for rack mounting, too.
XEGP can be installed into a rack by adding
two optional brackets to the left and right hand
side of the instrument.
HAZARD BY RACK
INSTALLATION
The front panel and its screws
are not designed to carry the
weight of the instrument!
Support the instrument, when
installed into a rack!
Disregarding may cause per-
sonal injury and damaged
equipment.
Alternative
DC Power inlet
and separate
fuse