Emerson Process Management GmbH & Co. OHG
9-3
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
9
Modbus Commands
9.3 List of Parameters and Registers - Sorted by Tag Name
9.3 List of Parameters and Registers - Sorted by Tag Name
Note!
The client access column in this list provides information about the read only (RO) or read/write
(R/W) access restrictions for each parameter.
Tag Name
Address
Data
Type
Client
Access
Description
Modicon
Daniel
Control.Acknowledge.AllStates
2091
2091
Boolean R/W
1=Acknowledge device's states; 0=no
effect
Control.Acknowledge.Failure
2092
2092
Boolean R/W
1=Acknowledge device's Namur Failure
alarms; 0=no effect
Control.Acknowledge.FctChecks
2095
2095
Boolean R/W
1=Acknowledge device's Namur Fct-
Check alarms; 0=no effect
Control.Acknowledge.LevelAlarms
2096
2096
Boolean R/W
1=Acknowledge device's level alarms;
0=no effect
Control.Acknowledge.MaintRequests
2094
2094
Boolean R/W
1=Acknowledge device's Namur Main-
tRequ alarms; 0=no effect
Control.Acknowledge.OffSpecs
2093
2093
Boolean R/W
1=Acknowledge device's Namur Off-
spec alarms; 0=no effect
Control.ApplyGas.PumpState1
2081
2081
Boolean R/W
Pump1 state (0=Off 1=On)
Control.ApplyGas.PumpState2
2082
2082
Boolean R/W
Pump1 state (0=Off 1=On)
Control.ApplyGas.SampleValve1
2051
2051
Boolean R/W
0=close all valves; 1=open sample
valve comp1
Control.ApplyGas.SampleValve2
2052
2052
Boolean R/W
0=close all valves; 1=open sample
valve comp2
Control.ApplyGas.SampleValve3
2053
2053
Boolean R/W
0=close all valves; 1=open sample
valve comp3
Control.ApplyGas.SampleValve4
2054
2054
Boolean R/W
0=close all valves; 1=open sample
valve comp4
Control.ApplyGas.SampleValve5
2055
2055
Boolean R/W
0=close all valves; 1=open sample
valve comp5
Control.ApplyGas.Span1Valve1
2061
2061
Boolean R/W
0=open sample valve; 1=open span1
valve comp1
Control.ApplyGas.Span1Valve2
2065
2065
Boolean R/W
0=open sample valve; 1=open span1
valve comp2
Control.ApplyGas.Span1Valve3
2069
2069
Boolean R/W
0=open sample valve; 1=open span1
valve comp3
Control.ApplyGas.Span1Valve4
2073
2073
Boolean R/W
0=open sample valve; 1=open span1
valve comp4
Control.ApplyGas.Span1Valve5
2077
2077
Boolean R/W
0=open sample valve; 1=open span1
valve comp5
Control.ApplyGas.Span2Valve1
2062
2062
Boolean R/W
0=open sample valve; 1=open span2
valve comp1
Control.ApplyGas.Span2Valve2
2066
2066
Boolean R/W
0=open sample valve; 1=open span2
valve comp2
Control.ApplyGas.Span2Valve3
2070
2070
Boolean R/W
0=open sample valve; 1=open span2
valve comp3