Emerson Process Management GmbH & Co. OHG
2-4
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
2.1 Common Technical Data
Optional interfaces for all models
Digital I/O board
7 or 14 digital inputs
(X-STREAM XEGK: max. 7 in-
puts)
electrical
specification
max. 30 V, internally limited to 2.3 mA
HIGH: min. 4 V; LOW: max. 3 V
common GND
function
Each input can be configured to any of the
functions listed by
Open valve
Activate sample gas pump
Zero calibrate all channels
Span calibrate all channels
Zero and span calibrate all channels
Abort calibration
9 or 18 additional relay outputs
(X-STREAM XEGK: max. 9 add.
outputs)
electrical
specification
Dry relay change-over contacts can be
used as NO or NC
max. load. 30 V; 1 A; 30 W resistive
function
Each output can be configured to provide
any of the functions listed by
.
These functions include, but are not limited
to
NAMUR NE 107 status signals 'Failure',
'Maintenance request', 'Out of specifica
-
tion', 'Function check' (these signals are
automatically configured Fail Safe),
concentration alarms (can manually be
configured Fail Safe),
control signals for external valves or
pumps,
and many more
Summary of Contents for Rosemount X-Stream Series