7
Maintenance & Procedures
7.2 General Maintenance Information
Tests
Component
Interval
Tubing, flexible
Leak Test
Once a year
Pressure sensor, pressure switch
Valve block
Pump diaphragm
Leak Test
After 4000 hrs of operation
(approx. 1/2 year, if continuously
operating)
Capillars
Pressure drop
Once a year
Flame arrestors
Pressure drop
See instructions in separate ma-
nual
Infallible containments
Several
See instructions in separate ma-
nual
RAW measuring values
Verify counts for zero
gases
(decreasing counts
may indicate conta-
mination of optical
components)
Monthly, then quarterly
Acceptable values:
photometer quotient: 1.0 ± 0.1
NO, N
2
O quotient: 1.0 ± 0.2
pO
2
, eO
2
, TC: 0 ± 100,000 counts
(for zero gas N
2
)
Replace Components Regularly
Component
Interval
Electrochemical oxygen cell
Depending on output signal (see details later in this sec-
tion)
Electrochemical TRACE oxygen cell Every six months
Filter, internal
Once a year, at least when contaminated
Filter, external
Several times a year, depending on process conditions