Emerson Process Management GmbH & Co. OHG
7-20
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
7.4.2 Manual Calibration
When finished, press
left
several times to re-
turn to
SELECT COMPONENT
(multi channel
analyzer only), to perform a span calibration
for another channel,
or
press
home
to return to the MEASUREMENT
SCREEN
, to finish with manual calibration
procedures.
7.4.2.2 Manual Span Calibration
Multi-channel unit:
Select the channel to be calibrated in
SELECT COMPONENT.
Before selecting any further line
make sure the required calibra-
tion gas is applied and flowing!
Span calibration offers the same options as
Zero calibration, so for a detailled description
Select
"Span calibration.."
Span calibration
Cancel!
Start!
Span gas
0.000 ppm
Concentration
0.000 ppm
Flow
0.00 l/min
Status..
Results..
Restore!
Zero calibration..
Span calibration..
Advanced calibration..
Apply gas..
Lock menus!
Acknowledgements
Pump 1:
Off
Pump 2:
Off
Ranges..