Emerson Process Management GmbH & Co. OHG
4-30
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
Feed the power cable through the cable
gland at the instrument´s right side and strip
the outer insulation. Strip the individual wires
and connect to the terminals (a label is loca
-
ted next to the terminals on the mains filter
housing).
Finally, tighten the outer dome nut to secure
the power cable.
Fig. 4-23:
Power
Supply Connections
4.6.2 Installation - X-STREAM XEXF Field Housings
Live L
Earth PE
Neutral N
Power supply
cable gland
ELECTRICAL SHOCK HAZARD
Verify the power supply at installation site meets the specification given on
the analyzer´s nameplate label, before installing the instrument!
Verify power cables are disconnected and/or instrument is de-energized
prior to working at the terminals!
Verify the power cord is layed with a distance of at least 1 cm (0.5“) to any
signal cable to ensure proper insulation from signal circuits!
Connecting the power cord
The power cord is connected to screw-type
terminals located inside the housing.
Accepted wire gauge:
0.2…4 mm
2
(AWG 24…AWG 12)
Recommended wire gauge
min. 1.5 mm² (AWG 15),
end sleeves not required
Skinning length:
8 mm (0.315")
Thread:
M3
Min. tightening torque:
0.5 Nm
(4.4 in.lb)