Emerson Process Management GmbH & Co. OHG
4-17
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
4
Installation
4.6.1 Installation - X-STREAM XEGK, X-STREAM XEGP
Fig. 4-12: Plug X5 - Analog Inputs
Analog inputs
Analog inputs are located on a 9-pole submin-D-connector (plug X5; Fig. 4-12) or on an optio
-
nal terminals adaptor XSTI (Fig. 4-13).
Pin
Signal
9
Input 2 high (+)
8
Input 2 low (-)
7
Input 1 low (-)
6
Input 1 high (+)
Signal
Pin
BR 2 fit a wire bridge here to apply analog
5
BR 2 signal in current mode to input 2 **)
4
unused
3
BR 1 fit a wire bridge here to apply analog
2
BR 1 signal in current mode to input 1 *)
1
{
}
{
}
*) alternatively set jumper P2 on electronics board XASI
**) alternatively set jumper P1 on electronics board XASI
2 analog inputs
electrical
specification
0–1 (10) V, software selectable; R
in
= 100 k
Ω
optional
(requires to fit wire bridges, see figures)
:
0
–
20 mA ; R
in
= 50
Ω
optically isolated from analyzer GND
protected against overload up to ±
15 V or ±
20 mA
Note!
Consider the installation notes in