Emerson Process Management GmbH & Co. OHG
7-26
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
7.4.3 Advanced Calibration
The analyzer immediately begins span
calibration(s), showing the CALIBRATION
STATUS SUMMARY screen.
Press
enter
in the first line to cancel the cur
-
rent calibration
"
Current action"
indicates, what currently is
carried out (
purging, zeroing, ready
)
"Action detail"
shows the current procedure,
or
Off
"
Current duration"
gives the remaining time
for the current procedure
"
Prev. duration"
shows the time elapsed since
start of procedure
"
Current step"
gives information about the
step currently carried out.
To see a detailled calibration status for a sin
-
gle channel, enter
CALIBRATION STATUS
SINGLE,
Calibration status summary
Cancel!
Calibration status single..
Current action
Ready
Action detail
Ch2
Current duration
0 s
Prev. duration
0 s
Current Step
0
Multi-channel unit:
Select the channel in SELECT COMPONENT.
to open the status screen with enhanced
calibration information about the current
channel (indicated in the uppermost display
line), including
remaining calibration time, currently measu
-
red concentration, setup of zero & span gas
concentrations and currently calibrated range.
The procedure has finished when '
Applied gas
'
shows
Sample gas
, or "
Current action
" in the
previous screen says
Ready
.
Press
home
to return to the MEASUREMENT
SCREEN.
Ch1
Calibration status single
Cancel!
Calibr.status
Ready
Remaining time
0 s
Concentration
0.000 ppm
Zero gas
0.000 ppm
Span gas
5000.000 ppm
Current range
Range 1
Applied gas
Zero gas