Emerson Process Management GmbH & Co. OHG
8-32
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
8.4 Troubleshooting on Components
Step 13:
Assemble the pump assy. Take care of
your marker (
step 4)
1.
Put the two upper plates under the
clamp (
steps 6 & 7 for reference).
2.
Put the white block and the
new
teflon gasket between the lower block
and the in-outlet plate.
Step 12:
Clean the white block.
Afterwards put in the new dia-phragms
and fix them with the new locking
springs.
1
2
Summary of Contents for Rosemount X-Stream Series