Installation
4.6.2 Installation - X-STREAM XEXF Field Housings
Fig. 4-18: Terminal Block X1 - Analog Signals and Relay Outputs 1-4
Note!
Consider the installation notes in
section
and the notes on installing cable glands
Pin
Signal
P2.1
Channel 1, (+) 4 (0) - 20 mA
P2.2
Channel 1, GND
P2.3
Channel 2, (+) 4 (0) - 20 mA
P2.4
Channel 2, GND
P2.5
Channel 3, (+) 4 (0) - 20 mA
P2.6
Channel 3, GND
P2.7
Channel 4, (+) 4 (0) - 20 mA
P2.8
Channel 4, GND
P2.9
Channel 5, (+) 4 (0) - 20 mA
P2.10
Channel 5, GND
P2.1
1
not used
P2.12
not used
P3.1
not used
P3.2
not used
P3.3
Output 1 (Failure), NC
P3.4
Output 1 (Failure), NO
P3.5
Output 1 (Failure), COM
P3.6
Output 2 (Maintenance Request), NC
P3.7
Output 2 (Maintenance Request), NO
P3.8
Output 2 (Maintenance Request), COM
P3.9
Output 3 (Out of Spec), NC
P3.10
Output 3 (Out of Spec), NO
P3.1
1
Output 3 (Out of Spec), COM
P3.12
Output 4 (Function check), NC
P4.1
Output 4 (Function check), NO
P4.2
Output 4 (Function check), COM
P4.3
not used
P4.4
P4.5
P4.6
P4.7
P4.8
P4.9
P4.10
P4.1
1
P4.12
Relay Outputs
**)
Analog Outputs
Serial Interface
*)
**)
Configuration of relay output termi
-
nals as per standard factory setting
(NAMUR status signals)
Analog signals
Relay outputs 1-4
Terminals for analog signals and relay outputs
1 - 4 are located on the outer left module
(terminal block X1; Fig. 4-18).
Specification of analog signal outputs:
4(0)–20 mA; burden: R
B
≤ 500 Ω
Specification of relay outputs 1-4:
max. 30 VDC, 1 A, 30 W
Note
Dry relay change-over contacts, can be used as
NO or NC.