6.2.3 Setup Menu
Firmware is the analyzer´s basic operation
software. This menu enables to update your
current analyzer firmware from an external
USB device, e.g. to add new features, etc.
6.2.3.8.5
Save-Load USB Firmware Update
Setup..
Save-Load..
USB Firmware Update..
Replace current firmware!
This will restart device!
Are you sure?
No!
Yes!
After selecting the firmware update procedu
-
re, a safety prompt appears: select "
Yes!"
to
start the procedure; "
No!"
cancels.
Firmware USB update procedure
•
The USB devices is checked for the directory
emerson
_
xe
/[
seral
numberIennummer
]/
fIrmware
/
Program
. If existant, the files within this directory are copied to the internal SD card. If not
existant, the USB device in the given order is scanned for the following directories to contain
valid firmware data
emerson
_
xe
/
uPgrade
/
fIrmware
/
Program
emerson
_
xe
/ [
serIal
number
]/
fIrmware
/
web
or
emerson
_
xe
/
uPgrade
/
fIrmware
/
web
.
The first found valid data are copied to the internal SD card.
• After the files are copied to the SD card, the analyzer reboots.
• After the reboot, the analyzer configuration files are copied to the SD card.
• Now the firmware files are copied to their final destination.
• Configuration files are restored from the SD card.
• All passwords are reset to factory defaults!
• A final analyzer reboot completes the update procedure.